If you like us, please share us on social media.
The latest UCD Hyperlibrary newsletter is now complete, check it out.

GeoWiki.png
ChemWiki: The Dynamic Chemistry E-textbook > Analytical Chemistry > Data Analysis > Instrument Calibration

MindTouch
Copyright (c) 2006-2014 MindTouch Inc.
http://mindtouch.com

This file and accompanying files are licensed under the MindTouch Master Subscription Agreement (MSA).

At any time, you shall not, directly or indirectly: (i) sublicense, resell, rent, lease, distribute, market, commercialize or otherwise transfer rights or usage to: (a) the Software, (b) any modified version or derivative work of the Software created by you or for you, or (c) MindTouch Open Source (which includes all non-supported versions of MindTouch-developed software), for any purpose including timesharing or service bureau purposes; (ii) remove or alter any copyright, trademark or proprietary notice in the Software; (iii) transfer, use or export the Software in violation of any applicable laws or regulations of any government or governmental agency; (iv) use or run on any of your hardware, or have deployed for use, any production version of MindTouch Open Source; (v) use any of the Support Services, Error corrections, Updates or Upgrades, for the MindTouch Open Source software or for any Server for which Support Services are not then purchased as provided hereunder; or (vi) reverse engineer, decompile or modify any encrypted or encoded portion of the Software.

A complete copy of the MSA is available at http://www.mindtouch.com/msa

Instrument Calibration

Calibration is the process of evaluating and adjusting the precision and accuracy of measurement equipment. Proper calibration of an instrument allows people to have a safe working environment and produce valid data for future reference.

Purpose of instrument calibration

Calibration refers to the act of evaluating and adjusting the precision and accuracy of measurement equipment. Instrument calibration is intended to eliminate or reduce bias in an instrument's readings over a range for all continuous values.

  • Precision is the degree to which repeated measurements under unchanged conditions show the same result
  • Accuracy is the degree of closeness of measurements of a quantity to its actual true value.

 

 This figure provides an example of precision and accuracy; it also shows the result after calibration.

 For this purpose, reference standards with known values for selected points covering the range of interest are measured with the instrument in question. Then a functional relationship is established between the values of the standards and the corresponding measurements. There are two basic situations:

Instruments which require correction for bias

The instrument reads in the same units as the reference standards. The purpose of the calibration is to identify and eliminate any bias in the instrument relative to the defined unit of measurement. For example, optical imaging systems that measure the width of lines on semiconductors read in micrometers, the unit of interest. Nonetheless, these instruments must be calibrated to values of reference standards if line width measurements across the industry are to agree with each other.

Instruments whose measurements act as surrogates for other measurements

The instrument reads in different units than the reference standards. The purpose of the calibration is to convert the instrument readings to the units of interest. An example is densitometer measurements that act as surrogates for measurements of radiation dosage. For this purpose, reference standards are irradiated at several dosage levels and then measured by radiometry. The same reference standards are measured by densitometer. The calibrated results of future densitometer readings on medical devices are the basis for deciding if the devices have been sterilized at the proper radiation level.

Basic steps for correcting the instrument for bias

The calibration method is the same for both situations stated above and requires the following basic steps:

  1. Selection of reference standards with known values to cover the range of interest.
  2. Measurements on the reference standards with the instrument to be calibrated.
  3. Functional relationship between the measured and known values of the reference standards (usually a least-squares fit to the data) called a Calibration Curve.
  4. Correction of all measurements by the inverse of the calibration curve.

How to reveal and avoid errors in the instruments

Some people mix up field check and calibration. Field check is when two instruments have the same reading; this does not mean they are calibrated; it may be that both instruments are wrong. Let's use thermometer as an example; if a thermometer always read .25 degree higher, this error can not be eliminated by taking averages, because this error is constant. The easiest way to determine if it is accurate and fix it is to send the thermometer to a calibration laboratory. Another way to reveal constant errors is to have one or more similar thermometers. One thermometer is used and then replaced by another thermometer. If readings are divided among two or more thermometers, inconsistencies among the thermometers will ultimately be revealed.

When do instruments need to be calibrated?

  • Before major critical measurements
    • Before any measurements that requires highly accurate data, send the instruments out for calibration and remain unused before the test.
  • After major critical measurements
    • Send the instrument for calibration after the test helps user decide whether the data obtained were reliable or not. Also, when using an instrument for a long time, the instrument's conditions will change.
  • After an event
    • The event here refers to any event that happens to the instrument. For example, when something hits the instrument or any kinds of accidents that might impact the instrument's accuracy. A safety check is also recommended.
  • When observations appear questionable
    • When you suspect the data's accuracy that is due to instrumental errors, send the instrument to calibrate.
  • Per requirements
    • Some experiments require calibration certificates. Check the requirements first before starting the experiment.
  • Indicated by manufacturer
    • Every instrument will need to be calibrated periodically to make sure it can function properly and safely. Manufacturers will indicate how often the instrument will need to be calibrated.

Why does calibration of instrument so important?

Let's say if you are going to publish a paper and you submitted the paper with data obtained from an uncalibrated instrument. What if someone repeated you experiment and find out that your result is wrong? This will hurt your reputation in the field, decrease the reliablitiy on your future works.

Another example, if you are going to work with a chemical that will explode when it gets in contact with air temperature above 50°C. So you adjust the room temperature before you start working, then check the temperature with an uncalibrated thermometer. If the termometer gives a lower temperature than the true temperature, then you will be working in an unsafe environment. This example may be unrealistic, but there are many chemical and substance out there that require accurate and precise measurements in order to provide others a safe working environemnt.

References

  1. Youden, W. J. Experimentation and Measurement. (1984)
  2. Skoog, DA. Holler, FJ. Crouch, SR. Principles of Instrumental Analysis, 6th Ed. (2007)
  3. Brown, Philip J. Measurement, Regression, and Calibration. Oxford statistical science series, 12. Oxford [England]: Clarendon Press (1993)

Contributors

You must to post a comment.
Last Modified
14:56, 15 Jul 2014

Tags

Classifications

(not set)
(not set)

Creative Commons License Unless otherwise noted, content in the UC Davis ChemWiki is licensed under a Creative Commons Attribution-Noncommercial-Share Alike 3.0 United States License. Permissions beyond the scope of this license may be available at copyright@ucdavis.edu. Questions and concerns can be directed toward Prof. Delmar Larsen (dlarsen@ucdavis.edu), Founder and Director. Terms of Use